Instructions for Use
Please send an email to info@floy.com to request a paper version.
Instructions for Use - BoneCT
The information provided by the manufacturer to inform the `User` of a device's `Intended Purpose` and proper use and of any `Precautions` to be taken.
Date of issue
2024-04-08
Device
[Medical Device: BoneCT]
Manufacturer
[Manufacturer: Floy]
Authorised Representative
N/A
Responsible Person
N/A
Clinical investigation
False
Clinical Benefits
[Clinical Benefit: Identify Patients at Risk]
Summary of Safety & Clinical Performance (SSCP)
N/A
Performance characteristics
[Device Data: BoneCT Device Data]
Residual risks
[
Risk Evaluation: Wrong Calibration to Unnecessary DEXA Scan;
Risk Evaluation: Unavailable Device To Unimproved Reading Evaluation;
Risk Evaluation: Credential Leak to PHI Deletion Evaluation;
]
Contra-indications
[
Contraindication: Metallic Implants;
Contraindication: Contrast-Enhanced;
]
Warnings
N/A
Disclaimers
[
Disclaimer: CT only;
Disclaimer: Native only;
Disclaimer: Spine only;
]
Precautions
[
Precaution: No children;
Precaution: No implants;
Precaution: No replacement;
Precaution: Second reader;
Precaution: Only Native;
]
---
Adverse Device Effects
N/A
Device Deficiencies
N/A
Limitation
- Inference time takes on average 2min and 52s but may take longer for larger series
Preparatory steps
- Add "Floy" as a new node within the established radiology network. This might require contacting your PACS provider.
Maintenance
- Regularly patch the underlying virtual machine provided to the manufacturer to maintain security.
Facility requirements and restrictions
- The IT network provided by the radiology office shall be secured and contain protection against malware.
Accessory and combination
[Accessory: Integration Engine]
Training requirements
No training necessary.
Qualification requirements
Users must be qualified healthcare professionals, e.g., radiology technologists, radiologists.
Consulting healthcare professionals
False
Serious incident reporting
Notice
Any serious incident that has occurred in relation to the device shall be reported to the manufacturer.
Any serious incident that has occurred in relation to the device shall be reported to the competent authority of the Member State in which the user and/or patient is established.
Procedure
Report serious incidents to the manufacturer by email info@floy.com or phone +4989244136090.
Special handling and storage
[Label: BoneCT Label [Special handling and storage]]
Device information
Installation statement
Installation must by performed by the manufacturer
Hardware requirements and restrictions
[Software Requirement Functional Capabilities: Hardware Environment]
Software requirements and restrictions
[Software Requirement Functional Capabilities: Software Environment]
Installation instructions
Install Floy Edge Relay on virtual machine provided by radiology clinic and connect securely with Floy.
Installation verification
Send a study to Floy and observe whether a Floy result is returned as attached DICOM series to the original study.
Critical dependencies
Requires connection with the [Accessory: Integration Engine].
Configuration requirements
Set AET title and IP address at defined here: https://www.floy.com/setup
System interface requirements and restrictions
Send applicable DICOM study from original PACS to Floy Edge Relay via DICOM Network (as usual by "Send to modality").
Start-up procedure
No startup procedure.
Shutdown procedure
No shutdown procedure.
Operating instructions
- Medical device returns a self-explanatory DICOM PDF Report as secondary capture.
- Radiology customer must ensure a secure IT network.
Glossary
AI: Artifical Intelligence
Security options
Provide the PACS AE title to Floy for whitelisting.
Failure
In case of failure to maintain security, immediately contact the manufacturer. As this represents a second reader device which provides add-on information, the core clinical workflow or patient remains unaffected.
Decommissioning
Contact the manufacturer for uninstallation and decommissioning.
Messages
[Instructions for Use Message: Not Supported Scan]
Website